Application
Characteristics:
 Black,
highly light absorbent, low contrast
Measurement
Requirements:
 Measurement
over a 150mm x 100mm area
 Measurement
of 3mm diameter blind bores
(required tolerance
of 2µm)
 Measurement
of 0.1mm bore depth
 Measurement
of internal moulded bosses
 Diameters;
radii; angular measurement
 Inspection
of surface detail
 Measurement
data recording and manipulation
Hawk
Configuration:
 Hawk
3-axis motorized non-contact
measurement system
 200
x 150mm high accuracy motorized stage
 x2
and x5 objective lens magnification
(total magnification
x20 and x50 respectively)
 6-point
ring light and epi-illumination
(sub-stage illumination
standard)
 QC5000
3D Metrology Software
Specification
Details: (Hawk 3-osiowy motorized system)

200mm x 150mm high accuracy motorized stage provides stage accuracy
better than 2µm, with motorized 'rapid' point-to-point movement
for increased sample throughput. x20 magnification used for diameter
and radii measurement. x50 magnification used for measuring 3mm
blind bore giving high magnification within the field of view and
a depth of focus of 4µm. 4µm depth of focus allow for
accurate measurement in the Z-axis. Through the lens epi-illumination
is used to illuminate the internal bore, while substage illumination
is used for profile measurement. QC5000 3D Metrology Software provides
all the required geometric functions, statistical reporting, part
programming and motorized stage control.
|